Enregistreur de niveau d'eau - MX20L
A cost-effective and high-performance data logger with Bluetooth data transfer for continuous monitoring of water level and temperature.
Measures up to 4, 9, or 30 meters in a wide variety of underwater environments, in both fresh and salt water.
Requires a compatible mobile device or Windows computer and the HOBOconnect app.
The MX20L Water Level Logger gives you all the same features as the U20L and more!
Features
The MX20L comes with a magnetic activation key to reactivate the logger and enable its Bluetooth communication for setup and data upload. The key comes with a lanyard to prevent loss or drop in the field. Additional activation keys are available for purchase.
The MX20L Water Level Logger gives you all the same features as the U20L and more!
Features
- The self-contained, non-vented design allows for easy deployment
- Ideal for use in fresh and salt water environments, including wells, streams, lakes, wetlands, and tidal areas
- Depth measurement options up to 4, 9, or 30 meters depending on the model
- Durable, freeze-resistant ceramic pressure sensor
- Simplified data transfer via Bluetooth to the free HOBOconnect app
- The HOBOconnect app allows easy retrieval and viewing of water level data
- Extended battery life of up to 10 years (power-save mode requires a magnetic key fob to enable Bluetooth for data upload)
The MX20L comes with a magnetic activation key to reactivate the logger and enable its Bluetooth communication for setup and data upload. The key comes with a lanyard to prevent loss or drop in the field. Additional activation keys are available for purchase.
This logger records absolute pressure, which is then converted into a water level measurement by our HOBOconnect software. To compensate for barometric pressure fluctuations, you can use another HOBO MX20L water level logger as a barometric reference.
Pressure (Absolute) and Water Level Measurements MX20L-01
- Operating Range
Pressure (Absolute) and Water Level Measurements MX20L-01
- Operating Range
0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) water depth at sea level, or 0 to 12 m (0 to 40 ft) water at 3,000 m (10,000 ft) elevation
- Factory Calibrated Range
69 to 207 kPa, 0° to 40°C
10 to 30 psia, 32° to 104°F
- Burst Pressure
310 kPa or 18 m depth
45 psia or 60 ft depth
- Water Level Accuracy*
- Typical Error: ± 0.1% FS, 1.0 cm (0.03 ft) water
- Maximum Error: ± 0.2% FS, 2.0 cm (0.06 ft) of water
- Raw Pressure Accuracy**
±0.3% FS, 0.62 kPa (0.09 psi) Maximum Error
- Resolution
- Factory Calibrated Range
69 to 207 kPa, 0° to 40°C
10 to 30 psia, 32° to 104°F
- Burst Pressure
310 kPa or 18 m depth
45 psia or 60 ft depth
- Water Level Accuracy*
- Typical Error: ± 0.1% FS, 1.0 cm (0.03 ft) water
- Maximum Error: ± 0.2% FS, 2.0 cm (0.06 ft) of water
- Raw Pressure Accuracy**
±0.3% FS, 0.62 kPa (0.09 psi) Maximum Error
- Resolution
< 0.02 kPa Water
0.007 ft of water
- Pressure Response Time (90%)***
0.007 ft of water
- Pressure Response Time (90%)***
< 1 second at stable temperature; Measurement accuracy also depends on temperature response time
MX20L-02 Pressure (Absolute) and Water Level Measurements
- Operating Range
0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) water depth at sea level, or 0 to 33.6 m (0 to 111 ft) water at 3,000 m (10,000 ft) elevation
- Factory Calibrated Range
69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure
500 kPa (72.5 psia) or 40.8 m (134 ft) depth
- Water Level Accuracy*
- Typical Error: ±0.1% FS, 3.0 cm (0.1 ft) water
- Error Maximum: ±0.2% FS, 6.0 cm (0.2 ft) of water
- Raw Pressure Accuracy**
±0.3% FS, 1.20 kPa (0.17 psi) Maximum Error
- Resolution
< 0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) of water
- Pressure Response Time (90%)***
< 1 second at stable temperature; measurement accuracy also depends on temperature response time
- Factory Calibrated Range
69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure
500 kPa (72.5 psia) or 40.8 m (134 ft) depth
- Water Level Accuracy*
- Typical Error: ±0.1% FS, 3.0 cm (0.1 ft) water
- Error Maximum: ±0.2% FS, 6.0 cm (0.2 ft) of water
- Raw Pressure Accuracy**
±0.3% FS, 1.20 kPa (0.17 psi) Maximum Error
- Resolution
< 0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) of water
- Pressure Response Time (90%)***
< 1 second at stable temperature; measurement accuracy also depends on temperature response time
MX20L-04 Pressure (Absolute) and Water Level Measurements
- Operating Range
0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) water depth at sea level, or 0 to 7 m (0 to 23 ft) water at 3,000 m (10,000 ft) elevation
- Factory Calibrated Range
69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure
310 kPa (45 psia) or 18 m (60 ft) depth
- Water Level Accuracy*
- Typical Error: ± 0.1% FS, 0.4 cm (0.013 ft) Water
- Maximum Error: ± 0.2% FS, 0.8 cm (0.026 ft) of water
- Raw Pressure Accuracy**
±0.3% FS, 0.43 kPa (0.063 psi) Maximum Error
- Resolution
< 0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) of water
- Pressure Response Time (90%)***
< 1 second at stable temperature; Measurement accuracy also depends on temperature response time
- Operating Range
0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) water depth at sea level, or 0 to 7 m (0 to 23 ft) water at 3,000 m (10,000 ft) elevation
- Factory Calibrated Range
69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure
310 kPa (45 psia) or 18 m (60 ft) depth
- Water Level Accuracy*
- Typical Error: ± 0.1% FS, 0.4 cm (0.013 ft) Water
- Maximum Error: ± 0.2% FS, 0.8 cm (0.026 ft) of water
- Raw Pressure Accuracy**
±0.3% FS, 0.43 kPa (0.063 psi) Maximum Error
- Resolution
< 0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) of water
- Pressure Response Time (90%)***
< 1 second at stable temperature; Measurement accuracy also depends on temperature response time
Temperature Measurements (All Models)
- Operating Range
-20° to 50°C
-4° to 122°F
- Accuracy
±0.2°C from 0° to 50°C
±0.36°F from 0° to 50°C
- Resolution
0.10°C at 25°C
0.18°F at 77°F
- Response Time (90%)
10 minutes in water (typical)
- Stability (Drift)
0.1°C per year
0.18°F per year
- Operating Range
-20° to 50°C
-4° to 122°F
- Accuracy
±0.2°C from 0° to 50°C
±0.36°F from 0° to 50°C
- Resolution
0.10°C at 25°C
0.18°F at 77°F
- Response Time (90%)
10 minutes in water (typical)
- Stability (Drift)
0.1°C per year
0.18°F per year
Logger
- Wireless Standard
- Bluetooth 5 LE
- Transmission Range
- Line of Sight Approximately 30.5 m (100 ft)
- Radio Power
1 mW (0 dBm)
- Real Time Clock
±1 minute per month from 0° to 50°C (32° to 122°F)
- Battery
2/3 AA, 3.6 volt lithium, factory replaceable
- Battery Life (Typical Use)
3 years with 1 minute recording and Bluetooth always on
7 years with 1 minute recording, Bluetooth off and LED off
10 years with 5 minute recording, Bluetooth off and LED off
- Battery Indication
Battery status shown in HOBOconnect app (approximate)
- Memory (non-volatile)
Approximately 400,000 sets of pressure and temperature measurements
- Time Data upload time (with Bluetooth 5+ devices)
Approx. 3.5 minutes; 100,000 measurements < 1 minute
- Weight
Approx. 154 g (5.43 oz) in air
Approx. 53.9 g (1.9 oz) in fresh water
- Dimensions
3.18 cm (1.25 in) diameter, 15.24 cm (6.0 in) length; 6.3 mm (0.25 in) diameter mounting hole
- Wireless Standard
- Bluetooth 5 LE
- Transmission Range
- Line of Sight Approximately 30.5 m (100 ft)
- Radio Power
1 mW (0 dBm)
- Real Time Clock
±1 minute per month from 0° to 50°C (32° to 122°F)
- Battery
2/3 AA, 3.6 volt lithium, factory replaceable
- Battery Life (Typical Use)
3 years with 1 minute recording and Bluetooth always on
7 years with 1 minute recording, Bluetooth off and LED off
10 years with 5 minute recording, Bluetooth off and LED off
- Battery Indication
Battery status shown in HOBOconnect app (approximate)
- Memory (non-volatile)
Approximately 400,000 sets of pressure and temperature measurements
- Time Data upload time (with Bluetooth 5+ devices)
Approx. 3.5 minutes; 100,000 measurements < 1 minute
- Weight
Approx. 154 g (5.43 oz) in air
Approx. 53.9 g (1.9 oz) in fresh water
- Dimensions
3.18 cm (1.25 in) diameter, 15.24 cm (6.0 in) length; 6.3 mm (0.25 in) diameter mounting hole
- Wetted Materials: Polypropylene housing and cord; Viton and Buna-N O-rings; ceramic sensor in an acetal tip; stainless steel screws suitable for salt water
- Logging Modes: Fixed interval (normal or statistical), burst, or multiple intervals with up to 8 user-defined intervals and durations; logging intervals from 1 second to 18 hours. See the HOBOconnect user guide for details.
- Start Modes: Immediate, next interval, date and time, button start (key fob or app)
- Stop Modes: Memory full, never stop (overwrite oldest data), date and time, or after the set logging period
- Pause: Pause and restart with the HOBOconnect app
- Alarms: Record high or low alarm events; Alarm status is displayed in the HOBOconnect app
- Environmental Rating: IP68
* Water Level Accuracy: With accurate reference water level measurement, known water density, accurate barometric compensation data, and a stable temperature environment.
** Raw Pressure Accuracy: The absolute accuracy of the pressure sensor includes all errors due to sensor drift, temperature, and hysteresis.
*** Temperature Variations: Soak the pressure sensor for 20 minutes in water to fully temperature compensate it. The maximum error due to rapid temperature variations is approximately 0.5%.
- Logging Modes: Fixed interval (normal or statistical), burst, or multiple intervals with up to 8 user-defined intervals and durations; logging intervals from 1 second to 18 hours. See the HOBOconnect user guide for details.
- Start Modes: Immediate, next interval, date and time, button start (key fob or app)
- Stop Modes: Memory full, never stop (overwrite oldest data), date and time, or after the set logging period
- Pause: Pause and restart with the HOBOconnect app
- Alarms: Record high or low alarm events; Alarm status is displayed in the HOBOconnect app
- Environmental Rating: IP68
* Water Level Accuracy: With accurate reference water level measurement, known water density, accurate barometric compensation data, and a stable temperature environment.
** Raw Pressure Accuracy: The absolute accuracy of the pressure sensor includes all errors due to sensor drift, temperature, and hysteresis.
*** Temperature Variations: Soak the pressure sensor for 20 minutes in water to fully temperature compensate it. The maximum error due to rapid temperature variations is approximately 0.5%.
Ref: MX20L